ΔT = qcal / Ccal  with  Ccal = Chardware + mwater cwater / 1000Calibrate reads it backwards from a known energy density  ·  Measure reads it forwards: qrxn = −qcal, then ΔUg = qrxn / m, then ΔUm = ΔUg M, then ΔHm = ΔUm + ΔngasRT  ·  with cwater = 4.184 J/(g·°C), R = 8.314 J/(mol·K) and T = 298.15 K

A weighed sample, a sealed steel vessel full of oxygen, and a thermometer in the water around it. This bomb calorimeter simulator is a free interactive lab that turns that one reading into an answer per gram and per mole. It opens on the calibration run: 1.00 g of benzoic acid in 2000 g of water, a rise of 2.6422 °C, and the sub-line so Ccal = q / ΔT = 10.000 kJ/°C.

Now look at where that constant came from: 8.368 kJ/°C of water and 1.632 kJ/°C of steel, stirrer and thermometer, each on its own slider. Every other free tool on this topic hands you the constant and hides the two halves of it — and dropping the second half is the one mistake a plausibility check cannot catch.

Bomb Calorimeter Lab

A bomb is rigid, so no work is done and the heat it measures is the internal energy change, not the enthalpy change. The rise follows from Ccal = Chardware + mwatercwater, and the same three numbers give the calorimeter constant one way and the heat of combustion the other. The last step, ΔH = ΔU + ΔngasRT, is decidable from the formula — and for graphite it is exactly nothing.

Temperature rise  ΔT = qcal / Ccal
2.6422 °C
so Ccal = q / ΔT = 10.000 kJ/°C
Heat  qcal = -ΔUg m
qcal = 26.422 kJ
qrxn = -26.422 kJ
Per gram and per mole  ΔUm = ΔUg M
ΔU = -26.422 kJ/g
-3226.76 kJ/mol
Enthalpy from internal energy  ΔH = ΔU + ΔngasRT
ΔH = -3228.00 kJ/mol
Δngas = -0.5, correction = -1.2394 kJ/mol
Constant Ccal 10.000 kJ/°C
Water term 8.368 kJ/°C
Hardware term 1.632 kJ/°C
Amount n = m / M 0.00818847 mol
Molar mass M 122.123 g/mol
Gas moles Δngas -0.5
Calorific value, gross (HHV) and net (LHV), kJ/g 26.43 gross; net not tabulated
Plausibility band band 0 — oxygen-bearing range: sugars, alcohols, biomass, coal
Sample mass m1.00 g
Water in the jacket mwater2000 g
Bomb hardware Chardware1.632 kJ/°C
Sample in the cup
Stage
Calibrate. Benzoic acid releases a known 26.422 kJ/g at constant volume, so 1.00 g of it gives qcal = 26.422 kJ. The thermometer rises 2.6422 °C, and q / ΔT therefore fixes this assembly's constant at Ccal = 10.000 kJ/°C — of which the water jacket carries 83.68 per cent and the steel and fittings the rest.
A bomb measures ΔU, not ΔH: the vessel is rigid, so the heat is the internal energy change and the enthalpy step is the labelled fifth line — exactly nothing whenever Δngas = 0 · the calibrant has two legitimate energy densities, both labelled here: 26.422 kJ/g at constant volume, which this bomb releases, and 26.43 kJ/g at constant pressure, which NIST tabulates — 0.04 % apart, and that gap is the point · NIST's own uncertainty is ±4 kJ/mol, 3.23 times the correction for the substance laboratories calibrate with · drag the hardware term to its floor for the error two ranking calculators are stuck in: a water-only 8.368 kJ/°C reads 16.3 % low and still looks plausible, so the band catches the impossible and misses this · the answer is the gross value, because the bomb's water charge keeps the combustion water liquid; never hand it to a boiler calculation that wants net · the drawn level and disc are schematic — they track their sliders, the disc radius as the cube root of the mass, but neither is to scale · Ccal belongs to one assembly on one day · a standard determination also corrects for the fuse wire, for the acids made from nitrogen and sulfur in the sample, and for the ignition energy, none of them modelled · this is the steady state, and the forward model knows each sample's tabulated energy density rather than discovering it.

Load a real machine

Each button presses the lab’s own stage button, then its own sample button, then writes its three sliders, so every press lands on a known state. Every number you then see is the lab’s own, and the line underneath quotes the cards straight back to you.

The first is the calibration run every real determination starts with. The second and third are the two corrections that bound the answer. The last two are the pair this page is built around: the apparatus changes, the reading changes, and the answer does not.

Pick a run above, or drag the sliders yourself.

What Is the Bomb Calorimeter Simulator?

The bomb calorimeter simulator is a free interactive physics lab that runs in your browser, with nothing to install and no sign-up. On the left it draws the apparatus cut away — the dewar, the water jacket, the stainless-steel bomb submerged in it, the cup holding the weighed sample, the oxygen fill at about 30 atm, the stirrer, the ignition leads and a thermometer marked at the start and finish readings. On the right it draws a linear energy-density ladder in kJ/g carrying published reference values, with a marker at whatever the current burn gives.

Three sliders and six buttons drive it. Sample mass runs 0.20 to 1.50 g in steps of 0.01 and opens at 1.00 g; Water in the jacket runs 1500 to 2500 g in steps of 10 and opens at 2000 g; Bomb hardware runs 1.000 to 3.000 kJ/°C in steps of 0.001 and opens at 1.632 kJ/°C. Four buttons choose what is in the cup — benzoic acid, graphite, ethanol or propan-1-ol — and two more choose the stage, Calibrate or Measure.

Say plainly what this is: a forward model. The lab knows each sample's tabulated energy density and works out the rise that would follow, so the Measure stage is showing you the arithmetic a student does rather than discovering an unknown. Nothing on the screen is a measurement. It is honest about that because the whole value of the panel is that you can check every step of the chain against a figure you can look up.

Four cards carry the chain. The headline is the temperature rise to four significant figures, with the constant it came from underneath; then the heat, positive into the assembly, with the reaction's own negative figure beneath it; then per gram and per mole; then enthalpy from internal energy. At the opening setting they read 2.6422 °C, qcal = 26.422 kJ, qrxn = -26.422 kJ, ΔU = -26.422 kJ/g with -3226.76 kJ/mol, and ΔH = -3228.00 kJ/mol.

The quantity this lab exists to expose is the calorimeter constant, and here it is two sliders and a grid cell. At the opening setting the grid reads 10.000 kJ/°C, of which 8.368 kJ/°C is the water jacket and 1.632 kJ/°C is the steel, the stirrer and the thermometer. Drag the jacket to 2500 g and the constant becomes 12.092 kJ/°C and the rise falls to 2.1851 °C for the same gram of sample. Every other free tool on this topic hides that number inside the arithmetic.

Calibrate and Measure share one piece of arithmetic and swap which end of it is the answer. In Calibrate the cup is forced to benzoic acid, whose energy density is known, and the sub-line under the rise card reads so Ccal = q / ΔT = 10.000 kJ/°C. In Measure that constant is taken as settled and the sub-line flips to from Ccal = 10.000 kJ/°C, with the rise feeding the other three cards. The two buttons are coupled on purpose: pressing a sample other than benzoic acid while the stage is Calibrate moves the stage to Measure, because a calibration run on graphite is not a calibration.

Halving the sample halves the rise and leaves the answer alone, and that is the whole reason calibration works. Take the sample to 0.50 g and the rise falls to 1.3211 °C and the heat to qcal = 13.211 kJ, while the per-gram card still reads ΔU = -26.422 kJ/g and the per-mole line still reads -3226.76 kJ/mol. Enlarging the jacket does the same thing to the rise and leaves the same two figures untouched. The apparatus moves the reading; it does not move the substance.

The enthalpy card is where the step every textbook asks about is sized rather than asserted. With benzoic acid it reads Δngas = -0.5, correction = -1.2394 kJ/mol against a molar figure of -3226.76 kJ/mol. Press Graphite and the same sub-line reads Δngas = 0, so ΔH = ΔU exactly, with both cards on -393.50 kJ/mol — because burning carbon makes exactly as many moles of gas as it consumes, and that is decidable from the formula before any arithmetic starts.

A compact grid under the cards carries the constant and both of its terms, the amount n = m / M, the molar mass, the change in gaseous moles, the gross and net calorific values with the gap between them as a percentage, and the plausibility band. For ethanol the gross-and-net cell reads 29.7 gross, 26.7 net; gap 10.10 per cent; for graphite it reads 32.8 gross, 32.808 net; no water produced, so the difference is rounding, which is the cleanest demonstration of what that gap is that the panel can give.

A sentence under the drawings writes the current case out in full rather than repeating the cards, and the ladder on the right places the answer against published figures: benzoic acid at 26.43, methane at 55.52, coal at 15.00 and damp peat at 6.00 kJ/g, with hydrogen at 141.79 drawn as a labelled off-scale arrow because putting it on the axis would squash everything else into nothing.

What the lab leaves out, stated as omissions rather than as assumptions that happen to hold. There is no picture of the reading climbing to its final value, because the approach to a steady temperature belongs to a different topic and a different page. There are four samples and no box to type an unknown one into, and the three corrections a standard-compliant determination makes — the fuse wire, the acids formed from nitrogen and sulfur, the ignition energy — are named on the page and given no numeric value anywhere.

All four samples sit in the lowest of the published bands, at or below 33.6 kJ/g, so the higher bands on the ladder are bounds to read against rather than states these buttons can reach. Nothing here is measured: every digit is arithmetic on three slider positions and two button presses, so verify any figure against your own source before you use it.

What you can change in the bomb calorimeter simulator, and what each control moves
ControlRangeStep, default and what it moves
Sample mass0.20 to 1.50 gsteps of 0.01 g, opening on 1.00, where the read-back beside the slider says "1.00 g" and the spoken value repeats it. This is the control that moves the reading without moving the answer: at 0.20 g the rise card reads 0.5284 degrees Celsius and at 1.50 g it reads 3.9633, while the per-gram card holds at -26.422 kJ/g in every position and the per-mole line at -3226.76 kJ/mol. The published practical range is about 1 to 1.5 g, so the slider brackets it at both ends
Water in the jacket1500 to 2500 gsteps of 10 g, opening on 2000, which is the jacket volume the fetched apparatus description gives and the figure that makes the water term exactly 8.368 kJ per degree Celsius. It is one of the two halves of the calorimeter constant: at 1500 g the grid reads a constant of 7.908 and a rise of 3.3412 degrees, at 2500 g a constant of 12.092 and a rise of 2.1851. More water means a smaller rise for the same burn, and the same answer in kJ/g at the end of it
Bomb hardware1.000 to 3.000 kJ/°Csteps of 0.001, opening on 1.632, which is the steel, the stirrer and the thermometer and which corresponds to about 3.9 kg of stainless steel. It is the other half of the constant, and the half every ranking calculator on this topic omits. At its floor the constant is 9.368 and the rise 2.8205 degrees; at its ceiling, 11.368 and 2.3243. The slider cannot reach zero, so the water-only mistake is reachable on the calculator but not on this panel
Sample in the cupfour buttonslabelled Benzoic, Graphite, Ethanol and Propanol, with Benzoic pressed when the lab opens. Each swaps the tabulated energy density, the molar mass, the formula and the change in gaseous moles together, so the whole chain moves at once: one gram gives a rise of 2.6422, 3.2762, 2.9619 or 3.3551 degrees Celsius and a molar answer of -3228.00, -393.50, -1367.00 or -2020.00 kJ/mol. Graphite is the one with no correction to make. There is no fifth sample and no way to type one in
Stagetwo buttonslabelled Calibrate and Measure, with Calibrate pressed at boot. Calibrate forces benzoic acid into the cup and reads the chain backwards, so the sub-line under the rise card gives the calorimeter constant from a known energy density; Measure takes that constant as settled and reads the chain forwards to kJ per gram and kJ per mole. The two groups are coupled: pressing any sample other than benzoic acid while the stage is Calibrate moves the stage to Measure, because a calibration run on graphite is not a calibration

How to use the bomb calorimeter equation simulator

  1. Read the four cards before you touch anything. The lab opens in Calibrate with 1.00 g of benzoic acid: Temperature rise 2.6422 °C, with so Ccal = q / ΔT = 10.000 kJ/°C under it; Heat qcal = 26.422 kJ over qrxn = -26.422 kJ; Per gram and per mole ΔU = -26.422 kJ/g over -3226.76 kJ/mol; and Enthalpy ΔH = -3228.00 kJ/mol.
  2. Read the grid underneath, because the headline number is built there. Constant 10.000 kJ/°C is Water term 8.368 kJ/°C plus Hardware term 1.632 kJ/°C, and the water half is the mass on the slider times the specific heat capacity of water, 4.184 J/(g·°C) — the figure the specific heat capacity of water guide settles on and explains the variants of.
  3. Drag the sample mass and watch which cards move. From 0.20 to 1.50 g the rise runs 0.5284 to 3.9633 °C and the heat card runs qcal = 5.284 kJ to 39.633 kJ. The per-gram and per-mole cards do not move at all. That is the first thing worth seeing on this panel.
  4. Press Half the sample, half the rise, then Bigger jacket, smaller rise. One changes the sample and one changes the apparatus; both change the rise, to 1.3211 and 2.1851 °C; and both leave ΔU = -26.422 kJ/g exactly where it was. The second also moves the constant to 12.092 kJ/°C, which is the water term going to 10.460.
  5. Drag the hardware slider to its floor and read the constant, not the rise. At 1.000 kJ/°C the constant reads 9.368 kJ/°C and the rise 2.8205 °C; at 3.000 they read 11.368 kJ/°C and 2.3243 °C. The slider stops at 1.000 and cannot reach zero, so the panel will not let you make the water-only mistake by accident — the bomb calorimeter calculator is where you can type it in and see what it costs.
  6. Press Graphite and read the bottom card only. The sub-line changes from a correction with a size to Δngas = 0, so ΔH = ΔU exactly, and both molar figures read -393.50 kJ/mol. Nothing was neglected and nothing was rounded away: burning carbon makes exactly as many moles of gas as it consumes.
  7. Press Ethanol and read the wide grid cell. It reads 29.7 gross, 26.7 net; gap 10.10 per cent, and ΔH = -1367.00 kJ/mol above it. The gap is the energy given up when the water the burn made condenses inside the vessel, which is the quantity the latent heat guide is about.
  8. Press Propanol for the fourth sample and the largest correction here. One gram gives 3.3551 °C, ΔU = -33.551 kJ/g and ΔH = -2020.00 kJ/mol, with Δngas = -1.5, correction = -3.7182 kJ/mol. Its gross-and-net cell reads 33.6 gross, 30.68 net; gap 8.69 per cent.
  9. Try to go back to Calibrate with graphite in the cup. You cannot, and the panel is right to stop you: pressing a sample other than benzoic acid while the stage is Calibrate moves the stage to Measure, because a calibration run on an unknown is not a calibration. The two button groups are coupled on purpose.
  10. Read the sentence under the drawings last. It writes the current case out in full rather than repeating the cards, and in Calibrate it adds the one figure the grid does not print: how much of the constant the water is carrying. At the opening setting that is 83.68 per cent; with the hardware slider at 3.000 it falls to 73.61 per cent.

Dragging is the right way to find out which readings move together and a poor way to land on a particular state, because the hardware slider alone has two thousand positions. The five preset buttons above exist for that, and the bomb calorimeter equation guide is where the subject itself is argued from scratch, with the chain derived step by step and the problems worked.

What this panel cannot do is start from numbers you measured yourself. It has no input box, four samples and no fifth, so a question that hands you a rise and a mass and asks for a molar mass — or hands you an enthalpy and asks what rise it would have produced — belongs to the bomb calorimeter calculator, which rearranges the same chain in five directions and prints the working for each.

Bomb calorimeter simulator on the Calibrate with benzoic acid preset, with two drawings on the left of a wide screen and the control panel on the right, under the lab's own heading Bomb Calorimeter Lab. The three sliders read Sample mass 1.00 g, Water in the jacket 2000 g and Bomb hardware 1.632 kilojoules per degree Celsius, with Benzoic pressed among the four sample buttons and Calibrate pressed among the two stage buttons. The four reading cards are Temperature rise, reading 2.6422 degrees Celsius, with the line so C sub cal equals q over the rise equals 10.000 kilojoules per degree Celsius beneath it; Heat, reading q sub cal equals 26.422 kilojoules with q sub rxn equals minus 26.422 kilojoules beneath it; Per gram and per mole, reading minus 26.422 kilojoules per gram with minus 3226.76 kilojoules per mole beneath it; and Enthalpy from internal energy, reading minus 3228.00 kilojoules per mole, with a change in gaseous moles of minus 0.5 and a correction of minus 1.2394 kilojoules per mole beneath it. The grid below gives Constant 10.000, Water term 8.368 and Hardware term 1.632 kilojoules per degree Celsius, Amount 0.00818847 mol, Molar mass 122.123 g/mol, Gas moles minus 0.5, Calorific value 26.43 gross with net not tabulated, and Plausibility band 0, the oxygen-bearing range of sugars, alcohols, biomass and coal. The left drawing is a cutaway titled benzoic acid C7H6O2, 1.00 g, labelled dewar, steel bomb, cup, water and thermometer, with the water filling roughly two thirds of the dewar and a gold disc for the sample sitting in the cup. Beside it a thermometer stem spanning 0 to 8 degrees above the start carries T underscore i 25.00 C at the foot, T underscore f 27.64 C at the top of its red column and the rise bracketed between them as dT 2.6422 C, and the caption along the foot reads 2000 g of water, C underscore cal equals 10.000 kJ per C. The right drawing is an energy-density ladder titled energy density, kJ/g, a linear axis from 0 to 60 ticked every 10, carrying methane 55.52, pentane 48.64, benzene 41.8, propan-1-ol 33.6, ethanol 29.7, benzoic acid 26.43, methanol 22.7, coal lignite 15.0 and peat damp 6.00, with the band above 55.52 shaded and labelled impossible unless hydrogen, hydrogen drawn above the top of the axis as an arrow labelled H2 141.79, off scale, and a gold marker alongside the benzoic acid tick. Its caption reads this burn 26.42; linear 0-60; H2 off scale, and above 55.52: H2 only. The sentence under the drawings begins Calibrate and says that benzoic acid releases a known 26.422 kilojoules per gram at constant volume, that 1.00 g of it therefore gives 26.422 kilojoules, that the thermometer rises 2.6422 degrees Celsius, and that the heat divided by the rise fixes this assembly's constant at 10.000 kilojoules per degree Celsius, of which the water jacket carries 83.68 per cent and the steel and fittings the rest.
The state every figure on this page is measured against: the Calibrate with benzoic acid preset. The rise card reads 2.6422 °C and its sub-line runs the chain backwards to Ccal = 10.000 kJ/°C, which is the only stage in which the apparatus is the answer. The ladder on the right marks the burn at the value its own caption gives, 26.42, which at this scale is indistinguishable from the published constant-pressure tick beside it at 26.43 — the same substance, measured two ways, each labelled. The canvas writes C_cal and T_f in ASCII, because a drawing surface has no markup to make a subscript with.

Worked example: change one thing at a time

Nothing in the four tables below was worked out here. Each row is a state the lab was put into through its own buttons and sliders and then read back from, cell by cell, and a reading is quoted whole rather than tidied: the heat card keeps its label, the per-gram card keeps its minus sign, and the gross-and-net cell keeps the clause that says why benzoic acid has no net figure.

Three kinds of cell are shortened to fit their column, and all three are declared here: the band cell to its number, the enthalpy column to its value without its sub-line, and the per-gram columns of the last two tables to the figure without the symbol the card prints in front of it.

One: two ways to change the reading, and one answer

The left-hand column is one gram of benzoic acid in Measure on the default apparatus. The middle column halves the sample and touches nothing else; the right-hand column leaves the sample alone and adds 500 g of water. Read the Temperature rise row and the Per gram row together, because that contrast is the entire argument for calibrating anything.

Every reading on the panel in three states: two of them change the apparatus or the sample, and none of them changes the answer
Reading on the panel One gram, standard jacket Half the sample Bigger jacket What moved, and why
Sample mass chip 1.00 g 0.50 g 1.00 g the only control the middle column touches
Jacket chip 2000 g 2000 g 2500 g the only control the right-hand column touches
Water term 8.368 kJ/°C 8.368 kJ/°C 10.460 kJ/°C the jacket’s share of the constant, and the only part a water slider can move
Constant Ccal 10.000 kJ/°C 10.000 kJ/°C 12.092 kJ/°C the hardware term stays 1.632 kJ/°C in all three
Temperature rise card 2.6422 °C 1.3211 °C 2.1851 °C all three differ. Less sample, or more to warm, means a smaller reading
Heat card qcal = 26.422 kJ qcal = 13.211 kJ qcal = 26.422 kJ halving the sample halves the heat; enlarging the jacket does not touch it
Heat of the reaction qrxn = -26.422 kJ qrxn = -13.211 kJ qrxn = -26.422 kJ the same magnitude with the sign the convention gives it
Amount burned 0.00818847 mol 0.00409423 mol 0.00818847 mol follows the mass, because the molar mass is a property of the substance
Per gram card ΔU = -26.422 kJ/g ΔU = -26.422 kJ/g ΔU = -26.422 kJ/g identical in all three. This is the answer, and the apparatus cannot move it
Per mole line -3226.76 kJ/mol -3226.76 kJ/mol -3226.76 kJ/mol identical, because it is the per-gram figure times a molar mass
Enthalpy card ΔH = -3228.00 kJ/mol ΔH = -3228.00 kJ/mol ΔH = -3228.00 kJ/mol identical, and equal to the published figure the sample table was built from
Plausibility band band 0 band 0 band 0 the same class of fuel, because it is the same substance

Three states, three different thermometer readings, and one number in the bottom half of the table that never moves. That is what a calorimeter constant buys you. Once the assembly has been calibrated, the reading it gives is a property of the apparatus and the sample together, while the figure you quote is a property of the substance alone.

It also sets the limit of what a reading on its own is worth. A rise of 2.1851 °C means -26.422 kJ/g in the right-hand column and would mean something else entirely in a different assembly, so a rise quoted without its constant is not a measurement of anything. That is why Calibrate is a stage rather than a footnote, and why the calculator version of this chain keeps the constant as one of its five solve directions.

Two: the sample mass from end to end

Benzoic acid in Measure, the jacket at 2000 g and the hardware at 1.632, with only the mass slider moving. Note that the 0.20 g row is the floor of this slider on this apparatus, not the smallest rise the lab can produce: the rise falls with the two apparatus terms as well, and the swept minimum across every combination is 0.3926 °C.

One calibrant in one assembly at seven sample masses, with both apparatus controls held still
Sample mass slider Temperature rise card Heat card Amount burned What the reading shows
0.20 g 0.5284 °C qcal = 5.284 kJ 0.00163769 mol the floor of this slider, and the smallest rise this apparatus gives
0.25 g 0.6606 °C qcal = 6.606 kJ 0.00204712 mol a quarter more sample, a quarter more rise, to four figures
0.50 g 1.3211 °C qcal = 13.211 kJ 0.00409423 mol the fourth preset: exactly half the rise of the 1.00 g row
0.75 g 1.9817 °C qcal = 19.817 kJ 0.00614135 mol three quarters, and the heat card is the rise with the point moved
1.00 g 2.6422 °C qcal = 26.422 kJ 0.00818847 mol the reference state, and the one the calibration preset uses
1.25 g 3.3028 °C qcal = 33.028 kJ 0.0102356 mol past the gram, and still exactly in step
1.50 g 3.9633 °C qcal = 39.633 kJ 0.0122827 mol the top of the slider, and the top of the published practical range

The rise and the heat are in strict proportion to the mass down that table, and the per-gram card holds at -26.422 kJ/g in every one of the seven rows. The heat card is the rise with the decimal point moved, which is the constant being exactly 10.000 kJ/°C rather than a coincidence; change the jacket and that tidy relation disappears while the per-gram figure stays.

Three: the two halves of the constant

Now the sample stays at one gram of benzoic acid and the two apparatus sliders move instead. The first five rows walk the jacket from end to end with the hardware fixed; the last three walk the hardware with the jacket fixed. The final column is the same string eight times, and that is the result rather than a formatting accident.

One gram of the calibrant in eight different assemblies, with the constant split into its two parts
Apparatus setting Water term Hardware term Constant Temperature rise card Per gram card
Jacket 1500 g 6.276 kJ/°C 1.632 kJ/°C 7.908 kJ/°C 3.3412 °C -26.422 kJ/g
Jacket 1750 g 7.322 kJ/°C 1.632 kJ/°C 8.954 kJ/°C 2.9509 °C -26.422 kJ/g
Jacket 2000 g 8.368 kJ/°C 1.632 kJ/°C 10.000 kJ/°C 2.6422 °C -26.422 kJ/g
Jacket 2250 g 9.414 kJ/°C 1.632 kJ/°C 11.046 kJ/°C 2.3920 °C -26.422 kJ/g
Jacket 2500 g 10.460 kJ/°C 1.632 kJ/°C 12.092 kJ/°C 2.1851 °C -26.422 kJ/g
Hardware 1.000 8.368 kJ/°C 1.000 kJ/°C 9.368 kJ/°C 2.8205 °C -26.422 kJ/g
Hardware 2.000 8.368 kJ/°C 2.000 kJ/°C 10.368 kJ/°C 2.5484 °C -26.422 kJ/g
Hardware 3.000 8.368 kJ/°C 3.000 kJ/°C 11.368 kJ/°C 2.3243 °C -26.422 kJ/g

The hardware term is between a tenth and rather more than a quarter of the constant across that table, and it is the term every ranking calculator on this topic leaves out. Leaving it out does not produce an absurd answer, which is exactly the problem: the reference run in this cluster gives 22.110 kJ/g against a true 26.422, which is 16.3 per cent low and still inside a plausible band.

The water half is the only part a reader can usually weigh directly, and it is the mass on the slider multiplied by the specific heat capacity of water. If that product is the step you want to practise on its own, the specific heat capacity guide owns it and the specific heat simulator puts it on its own screen, where the material and the mass are the controls instead of one half of a constant.

Four: the four samples, and the two corrections

One gram of each, in Measure, on the default apparatus. The last two columns are the payload of the whole topic: the change in gaseous moles is read off the molecular formula before any arithmetic starts, and the gross-and-net cell says whether the water the burn made is inside the figure.

What one gram of each sample does on the same assembly, with both corrections beside it
Sample in the cup Temperature rise card Per gram card Enthalpy card Gas moles Gross and net cell
Benzoic acid C7H6O2 2.6422 °C -26.422 kJ/g -3228.00 kJ/mol -0.5 26.43 gross; net not tabulated
Graphite C 3.2762 °C -32.762 kJ/g -393.50 kJ/mol 0 32.8 gross, 32.808 net; no water produced, so the difference is rounding
Ethanol C2H6O 2.9619 °C -29.619 kJ/g -1367.00 kJ/mol -1 29.7 gross, 26.7 net; gap 10.10 per cent
Propan-1-ol C3H8O 3.3551 °C -33.551 kJ/g -2020.00 kJ/mol -1.5 33.6 gross, 30.68 net; gap 8.69 per cent

The graphite row is the cleanest demonstration this panel can give, and it needs no arithmetic at all. Carbon makes no water, so gross and net must coincide, and the tabulated pair reads 32.8 against 32.808. Carbon also gains no moles of gas, so the enthalpy and the internal energy are the same figure, and the card says so ΔH = ΔU exactly rather than quoting a tiny correction.

Formula and symbol reference

The band at the head of this page carries two relations, and every reading on the panel is a factor in one of them or a consequence of them. The rise is the only quantity the lab computes forwards from a tabulated energy density; everything below it is that rise being carried back out again.

The symbol column below uses the panel’s own names, and the “in this lab” column gives each slider’s real end stops with the readings measured at those positions. Six of the rows are not controls at all: they are computed, fixed, or one of four values set by a sample button.

Every quantity this panel names, with the span it covers and a reading from each end
Symbol What it is SI unit In this lab
ΔT Temperature rise of the whole assembly. It is a difference, so degrees Celsius and kelvin are the same number here and nothing is added to convert °C the headline card, four significant figures. Across every combination of the three sliders and the four samples it runs from 0.3926 °C (benzoic acid, 0.20 g, 2500 g, 3.000) to 6.9168 °C (propan-1-ol, 1.50 g, 1500 g, 1.000), swept over the real grids rather than reasoned about. The all-minimum corner is not the floor: it reads 0.7263 °C.
Ccal Calorimeter constant, also called the energy equivalent: the energy the whole assembly absorbs for each degree it warms kJ/°C a grid cell, and the quantity this lab exists to expose. It is never a slider: it is built from the two below it. The panel reaches 7.276 kJ/°C at the bottom of both apparatus sliders and 13.460 kJ/°C at the top of both, and sits at 10.000 kJ/°C on the opening setting.
mwater Mass of water in the jacket. The first of the two things that absorb the heat g the second slider, 1500 to 2500 g in steps of 10, opening on 2000 g — the jacket volume the fetched apparatus description gives. It moves the water term from 6.276 to 10.460 kJ/°C and the rise from 3.3412 to 2.1851 °C for one gram of the calibrant.
Chardware Heat capacity of the steel bomb, the stirrer and the thermometer. The second of the two, and the one most calculations on this topic drop kJ/°C the third slider, 1.000 to 3.000 in steps of 0.001, opening on 1.632, which corresponds to about 3.9 kg of stainless steel. It cannot be dragged to zero, so the water-only mistake is not reachable from this panel; its floor still leaves the constant at 9.368 kJ/°C.
cwater Specific heat capacity of water, the number that turns a jacket mass into a heat capacity J/(g·°C) fixed at 4.184 and not exposed as a control. It is the thermochemical calorie by definition, and it is why 2000 g of water is exactly 8.368 kJ/°C. No slider on this panel changes it and no sample affects it.
m Mass of sample weighed into the cup g the first slider, 0.20 to 1.50 g in steps of 0.01, opening on 1.00 g, which brackets the published practical range at both ends. It is the one control that moves the rise and the heat together while leaving the per-gram and per-mole cards exactly where they were.
qcal Heat absorbed by the assembly. Positive, because the thermometer went up kJ the second card. Measured across the sliders from 5.284 kJ (0.20 g of the calibrant) to 50.327 kJ (1.50 g of propan-1-ol). It is always the sample’s energy density times its mass, never the constant times the displayed rise, because those two routes disagree in the last float on about a tenth of the reachable states.
qrxn Heat of the reaction. The negative of the heat above, by energy conservation kJ the sub-line of the second card, and the first place the sign appears. For a combustion it is always negative and it is always printed with its minus sign; there is no setting of these controls that makes it positive.
ΔUg Internal energy of combustion per gram. What a rigid vessel measures, because it does no pressure-volume work kJ/g the third card, and the quantity the apparatus cannot move. It takes exactly four values here, one per button: -26.422, -32.762, -29.619 and -33.551 kJ/g. The ladder beside the drawing plots the magnitude of this figure against published values.
n Amount of sample burned, the mass divided by the molar mass mol a grid cell, six significant figures. From 0.00163769 mol (0.20 g of the calibrant) to 0.124886 mol (1.50 g of graphite, whose molar mass is the smallest of the four). It is the only quantity on the panel that depends on both the slider and the button.
M Molar mass of whatever is in the cup g/mol a grid cell with exactly four values, because there are exactly four samples: 122.123, 12.011, 46.069 and 60.096 g/mol. None of them is a control, and the first is within 0.0014 per cent of the molar mass the fetched source states for benzoic acid.
ΔUm Internal energy of combustion per mole: the per-gram figure multiplied by the molar mass kJ/mol the sub-line of the third card, two decimals. -3226.76, -393.50, -1364.52 and -2016.28 kJ/mol for the four samples. Only the button changes it; no slider on the panel does.
Δngas Change in the number of moles of gas when the sample burns to carbon dioxide and liquid water. It may be negative, and it may be a half-integer none — a count a grid cell, read off the molecular formula rather than measured. -0.5, 0, -1 and -1.5 for the four samples. Zero is the interesting value, and it is the graphite column: no gas is gained or lost, so the last card has nothing to do.
ΔHm Enthalpy of combustion per mole: the internal energy plus the pressure-volume term for the gases kJ/mol the fourth card. -3228.00, -393.50, -1367.00 and -2020.00 kJ/mol. The correction that separates it from the card above runs -1.2394, 0, -2.4788 and -3.7182 kJ/mol, and the card prints both the size and the sign of it every time.
HHV and LHV Gross and net calorific values: with and without the energy released when the water the burn makes condenses kJ/g one wide grid cell, quoted as positive magnitudes and always labelled. 29.7 gross, 26.7 net; gap 10.10 per cent for ethanol; 33.6 gross, 30.68 net; gap 8.69 per cent for propan-1-ol; 32.8 gross, 32.808 net for graphite, where the difference is rounding. For benzoic acid the cell says net not tabulated rather than inventing one.

The division by a thousand in the constant is the conversion most often dropped by hand. The specific heat capacity of water is quoted in joules and the whole panel works in kilojoules, so a jacket of 2000 g is 8.368 kJ/°C and not 8368. Getting it wrong by a factor of a thousand puts the rise somewhere no thermometer reads, which at least makes it visible.

Two symbols in that table collide with other pages on this site and are worth separating before you substitute. The capital letter used here for the calorimeter constant is a heat capacity in kilojoules per degree, not the capacitance of the same name, and the lower-case one is the specific heat capacity of a substance, not a speed. Both conventions are standard; neither page is wrong, and no reading here should be carried into a formula from the other.

The physics of a constant-volume burn

The reason a bomb is made of thick steel is not safety alone. A rigid vessel cannot change volume, so the burning sample pushes nothing outwards and does no pressure-volume work — which means the energy it gives up is exactly the internal energy change, with no correction and no assumption. The first law of thermodynamics guide is where that step is set out properly, and the first law simulator is where the work term is a control you can take to zero yourself.

That one fact is why this apparatus measures what it measures. A reaction run in an open flask at atmospheric pressure does push the air back, and what you get there is an enthalpy change. The same reaction in the bomb gives an internal energy change instead, and the two differ by the work the gases would have done. On a pressure-volume plane a constant-volume path is a vertical line enclosing no area, which the pressure-volume diagram simulator draws directly.

The size of that difference is the last card on the panel, and the whole point is that it is decidable before you start. For a condensed fuel burning to carbon dioxide and liquid water, the change in gaseous moles depends on the hydrogen and oxygen counts alone and the carbon count cancels out, so you can read it off the formula. The grid prints it as -0.5, 0, -1 or -1.5 for the four samples, and the card turns it into a correction through the ideal gas relation.

And it is exactly zero whenever the hydrogen count is twice the oxygen count, which is the carbohydrate condition. A carbohydrate already carries its hydrogen and oxygen in the ratio they have in water, so burning it makes precisely as many moles of gas as it consumes. Graphite is the degenerate case of the same thing, which is why its two molar cards are the same number rather than two numbers that round alike.

The correction itself arrives through the relation between pressure, volume and amount for a gas — the ideal gas law guide is where that comes from — evaluated at 298.15 K with R = 8.314 J/(mol·K). That makes the whole term worth 2.4788 kJ for each mole of gas gained or lost, which is where the panel’s -1.2394, -2.4788 and -3.7182 kJ/mol come from.

Keep the size of it in proportion. For benzoic acid the correction is 0.0384 per cent of the internal energy figure, while the published uncertainty on the calibrant itself is 0.124 per cent — so the uncertainty on the standard substance is 3.23 times the size of the correction every textbook devotes a section to. That is not an argument for skipping the step. It is the measurement that tells you when the step can matter and when it cannot.

The second correction is physical rather than arithmetic. The vessel holds a little water of its own so that its atmosphere is saturated and everything the reaction makes ends up liquid, which means the energy given up as that water condenses is inside the reading. The answer is therefore a gross value, and the ladder on the right plots it against published gross values for the same reason.

The five parts of the two drawings, what each one paints and what it refuses to claim
Part of the drawing What it paints What it does not claim
the apparatus, left panel A cutaway titled with the sample and its mass — benzoic acid C7H6O2, 1.00 g — drawing seven labelled parts at a wide canvas: the water, the steel bomb, the cup, the dewar, the stirrer, the ignition leads and O2 at about 30 atm. The caption along its foot reads 2000 g of water, C_cal = 10.000 kJ/C none of it is to scale as an engineering drawing. The jacket level and the sample disc are driven by the two sliders so the picture cannot contradict the cards, but no dimension in the scene is a length you can measure off, and the number of labelled parts drops to four on a narrow canvas rather than overprinting
the thermometer A stem covering a fixed span of 0 to 8 C above the start, with the starting reading at T_i, the finish at T_f and the rise bracketed between them. At a narrow canvas the readings move into the caption strip instead, as T_f 27.64 C, dT = 2.6422 C there is no elapsed time on this drawing and there never will be, because the approach to a steady reading belongs to a different topic. The span is fixed rather than fitted to the current rise, and every rise these sliders can reach sits inside it — an assertion in the reference suite, not an assumption
the ladder, right panel A linear axis in kilojoules per gram from 0 to 60 with ticks every 10, carrying nine labelled published values at the width the figures on this page were captured at: benzoic acid 26.43, methane 55.52, pentane 48.64, propan-1-ol 33.6, ethanol 29.7, benzene 41.8, methanol 22.7, lignite coal 15.0 and damp peat 6.00. Its title degrades with the width, from energy density kJ/g, selected fetched values on the widest canvas to energy density, kJ/g on the figures below the axis is linear and the caption says so, because a logarithmic one over 6 to 142 would spend most of its length on empty space. The marker and the tick values share one scale, so the marker’s position against a named value is a real comparison rather than a decoration
hydrogen, and the shaded zone Hydrogen is drawn above the top of the axis as an explicit arrow labelled H2 141.79, off scale, and the band above 55.52 is shaded and labelled impossible unless hydrogen, with the caption repeating above 55.52: impossible unless hydrogen nothing is silently omitted and nothing is compressed to make room. Putting hydrogen on the axis would squash the 15 to 50 region the whole topic lives in; dropping it would hide the real ceiling. The shading is drawn exactly when its claim is stated in words, never one without the other
what the marker can reach A gold marker at the magnitude of the current per-gram figure. With these four buttons it only ever sits at 26.42, 32.76, 29.62 or 33.55 the higher bands are bounds to read against, not states these buttons can reach. All four samples sit in the lowest published band, at or below 33.6 kJ/g, so a reader who wants to see the marker in the hydrocarbon range has to take the figure to the calculator rather than expecting a button here to produce it
Bomb calorimeter simulator on the Graphite preset, the same apparatus with the sample button changed and the stage moved to Measure. The three sliders still read Sample mass 1.00 g, Water in the jacket 2000 g and Bomb hardware 1.632 kilojoules per degree Celsius, but Graphite is now pressed among the four sample buttons and Measure among the two stage buttons. The Temperature rise card reads 3.2762 degrees Celsius with from C sub cal equals 10.000 kilojoules per degree Celsius beneath it, the Heat card reads q sub cal equals 32.762 kilojoules over q sub rxn equals minus 32.762 kilojoules, the Per gram and per mole card reads minus 32.762 kilojoules per gram over minus 393.50 kilojoules per mole, and the Enthalpy card reads minus 393.50 kilojoules per mole, the same figure again, with the line the change in gaseous moles is 0, so the enthalpy change equals the internal energy change exactly beneath it. In the grid below, Molar mass has changed to 12.011 g per mol, Amount to 0.0832570 mol and Gas moles to 0, while Constant 10.000, Water term 8.368 and Hardware term 1.632 kilojoules per degree Celsius are unchanged, and the calorific value cell now reads 32.8 gross, 32.808 net, no water produced, so the difference is rounding. The left drawing is titled graphite C, 1.00 g, with the same dewar, steel bomb, cup, water and thermometer labels, and its thermometer now carries T underscore i 25.00 C, T underscore f 28.28 C and a bracket reading dT 3.2762 C, with the caption 2000 g of water, C underscore cal equals 10.000 kJ per C. On the energy-density ladder the gold marker has moved up to just below the propan-1-ol tick at 33.6 and the caption now reads this burn 32.76; linear 0-60; H2 off scale, while the shaded zone above 55.52 and the off-scale hydrogen arrow at 141.79 are unchanged. The sentence under the drawings begins Measure and says that with the constant already calibrated at 10.000 kilojoules per degree Celsius, 1.00 g of graphite C gives a 3.2762 degree rise, so the heat of the reaction is minus 32.762 kilojoules, which is minus 32.762 kilojoules per gram and minus 393.50 kilojoules per mole, and that because the change in gaseous moles is 0 there is nothing to add.
The Graphite: internal energy equals enthalpy preset, and the one state on this panel where the last card has nothing to do. Both molar figures read -393.50 kJ/mol and the sub-line says Δngas = 0, so ΔH = ΔU exactly. The gross-and-net cell makes the other correction vanish too, at 32.8 gross, 32.808 net, because carbon makes no water for the vessel to condense.

Where the bomb calorimeter simulator breaks down

Every state the three sliders and six buttons reach is answered, and the arithmetic is not where this panel can let you down. What it cannot be is a standard-compliant determination, and what it refuses to do is treat its own idealisations as facts. The first two items below matter more than the rest together.

The textbook equation is the constant times the rise, and a real determination is not
This is the hard limit. A standard-compliant measurement also corrects for the burning fuse wire, for the nitric and sulfuric acid formed from nitrogen and sulfur in the sample and found by titrating the residual liquid, and for the electrical ignition energy. All three are named in the fetched apparatus description and all three are codified: ASTM D5865 and ISO 1928 for solid fuels, ASTM D240 and D4809 for liquid hydrocarbons. This lab names them and attaches a number to none of them, because none was fetched. Verify against the standard before you use a figure for anything that matters.
The constant belongs to that assembly on that day
Nothing makes 10.000 kJ/°C a property of bomb calorimeters in general. The table above puts eight different constants on the same gram of the same substance, from 7.908 to 12.092 kJ/°C, simply by moving two sliders. A constant has to be re-determined for the assembly in front of you, and that is the entire reason the Calibrate stage exists rather than being a number printed in a manual.
It is a forward model, not a measurement
The lab already knows the answer. It holds each sample’s tabulated energy density and works out the rise that would follow, so the Measure stage is showing you the arithmetic a student does rather than discovering an unknown. That is honest and it is deliberate: the value of the panel is that every step can be checked against a figure you can look up, which a black box cannot offer. No digit on this screen is an observation.
The hardware slider cannot reach zero, so the commonest error is not reachable here
Two of the ranking calculators on this topic use the water term alone, with no hardware at all. The reference run in this cluster sizes that: a reader who correctly measures a rise of 2.6422 °C on a gram of the calibrant but applies 8.368 kJ/°C gets 22.110 kJ/g against a true 26.422, 16.3 per cent low and always low. This slider stops at 1.000, so the panel will not produce that state; the calculator will, if you type it.
The plausibility band catches the impossible and misses the merely wrong
The band compares a per-gram figure with four published bounds — 33.6, 48.64, 55.52 and 141.79 kJ/g — and the error above sails through it. So the band is a filter on the shape of an answer and not a proof of one. It is also worth knowing that all four samples here sit in the lowest band, at or below 33.6 kJ/g, so the higher bands are bounds the ladder draws rather than states these buttons can produce.
The answer is a gross value, and some calculations want net
Everything this panel prints includes the energy released when the water the sample made condenses, because the vessel is built to make that happen. An engine, a boiler or a fuel-economy figure usually wants the net value instead, which excludes it. The grid gives you both where a net figure is tabulated and says net not tabulated where one is not, and the gap is the quantity the latent heat guide explains and the latent heat simulator draws.
The enthalpy step assumes ideal gases and ignores the condensed phases
The correction treats every gaseous species as ideal and assumes the solids and liquids contribute no pressure-volume term of their own. Both approximations are good to well under the published uncertainty on the calibrant, so neither is the limiting error in anything this lab reports — but they are approximations, and the sign of the correction is as important as its size. A sample that gains moles of gas would push the enthalpy the other way.
Nitrogen and sulfur in the sample change the answer before the arithmetic does
A sample containing either forms acid in the residual liquid, and that energy is in the reading until the first correction above takes it out. The fetched table includes a nitrogen-bearing solid at 30.0 kJ/g, comfortably inside the band this lab would call plausible, which is precisely why a band check cannot substitute for a titration. None of the four samples here contains nitrogen or sulfur.
Nothing here happens over time
The panel reports the finished state and has no clock at all. How the reading approaches its final value, how long the stirring takes to even the jacket out, and what the vessel loses to the room while you wait are separate questions with separate formulas, and none of them is on this screen. The drawing shows the thermometer with its start and finish marked and the rise bracketed between them, and nothing is plotted against elapsed seconds anywhere on the page.
Four samples, three sliders, and no measurement anywhere
You choose a stage, choose a sample and set three controls, and the lab reports what they imply. There is no box to type a measured rise into, no fifth sample, no unit menu and no way to run the chain backwards from a molar answer to a mass. The sample table is assembled from fetched literature values, and one of them — the molar mass of the calibrant — is the only figure in the set checked directly against its source. Verify any figure against your own source before you use it.

Where bomb calorimetry is actually used

Calibrating an assembly before you believe a single reading from it
This is what the first preset is, and it is the question no calculator on this topic puts first. Burn a known mass of the standard substance, read the rise, and the sub-line hands you the constant: 1.00 g and 2.6422 °C give 10.000 kJ/°C. Burn half as much and it still gives 10.000 kJ/°C, from a rise of 1.3211 °C — which is the check that tells you the calibration is sound rather than a coincidence.
Turning one thermometer reading into a figure a textbook will accept
The four cards are the four steps, in order, and the panel never skips one. A rise of 2.9619 °C on a gram of ethanol becomes qcal = 29.619 kJ, then -29.619 kJ/g, then -1364.52 kJ/mol, then ΔH = -1367.00 kJ/mol. If you want that same chain with your own numbers in the boxes, the bomb calorimeter calculator is the page for it.
Deciding whether the enthalpy step is worth doing at all
Read the molecular formula, read the gas-moles cell, and you know before you start. It is 0 for graphite and for every carbohydrate, so the step is exactly nothing and the two molar cards coincide. It is -1.5 for propan-1-ol, the largest here, and the correction is still only -3.7182 kJ/mol on a figure of -2016.28. Knowing which case you are in is more useful than doing the arithmetic twice.
Telling a gross figure from a net one before you quote it
Fuel data is published both ways and the difference is not small: the panel shows 29.7 against 26.7 kJ/g for ethanol, a gap of 10.10 per cent. Anything to do with combustion efficiency, fuel comparison or energy content needs to know which it has. The quickest way to see what the gap is rather than how big it is, is to press Graphite and watch it disappear.
Sanity-checking a per-gram answer against published values
The ladder is there for exactly that. It carries nine fetched reference values from damp peat at 6.00 to methane at 55.52 kJ/g, with hydrogen drawn off-scale at 141.79, and the marker sits at your own figure on the same axis. A figure above 55.52 is impossible for anything but hydrogen; a figure above 33.6 says the sample cannot be the oxygen-bearing kind. That is a filter worth applying, and the section above says plainly what it will not catch.
Seeing why a different laboratory gets a different reading and the same answer
Two assemblies with different jackets and different steel give different rises for identical samples — 3.3412 °C against 2.1851 °C for the same gram here — and both report -26.422 kJ/g. That is the property that makes the technique reproducible between laboratories at all, and it is visible on this panel in two clicks rather than being asserted in a paragraph.
Knowing when this relation is the wrong tool
Two cases make it wrong rather than imprecise. If what you already have is an amount in moles and a molar enthalpy, and what you want is a heat, the chain runs the other way and the reaction enthalpy calculator is the right page. And if your vessel is open to the atmosphere rather than sealed, the heat you measure is an enthalpy change from the start and there is no gas-moles step to add.
Bomb calorimeter simulator on the Bigger jacket, smaller rise preset, the same gram of benzoic acid with only the water slider moved. The three sliders read Sample mass 1.00 g, Water in the jacket 2500 g and Bomb hardware 1.632 kilojoules per degree Celsius, with Benzoic pressed and the stage on Measure. The Temperature rise card reads 2.1851 degrees Celsius, lower than the 2.6422 of the 2000 g state, with from C sub cal equals 12.092 kilojoules per degree Celsius beneath it. The Heat card still reads q sub cal equals 26.422 kilojoules over q sub rxn equals minus 26.422 kilojoules, the Per gram and per mole card still reads minus 26.422 kilojoules per gram over minus 3226.76 kilojoules per mole, and the Enthalpy card still reads minus 3228.00 kilojoules per mole with a change in gaseous moles of minus 0.5 and a correction of minus 1.2394 kilojoules per mole. In the grid below, Constant has risen to 12.092 and Water term to 10.460 kilojoules per degree Celsius while Hardware term holds at 1.632, and Amount, Molar mass, Gas moles, the calorific value cell and the plausibility band are all unchanged. The left drawing is titled benzoic acid C7H6O2, 1.00 g, and the water level inside the dewar is visibly higher than in the 2000 g figures, reaching close to the top of the steel bomb; its thermometer carries T underscore i 25.00 C, T underscore f 27.19 C and a bracket reading dT 2.1851 C, and the caption along the foot now reads 2500 g of water, C underscore cal equals 12.092 kJ per C. On the energy-density ladder the gold marker sits exactly where it did in the calibration figure, alongside the benzoic acid tick at 26.43, and the caption still reads this burn 26.42; linear 0-60; H2 off scale. The sentence under the drawings begins Measure and says that with the constant already calibrated at 12.092 kilojoules per degree Celsius, 1.00 g of benzoic acid C7H6O2 gives a 2.1851 degree rise, so the heat of the reaction is minus 26.422 kilojoules, which is minus 26.422 kilojoules per gram and minus 3226.76 kilojoules per mole, and that adding the correction of minus 1.2394 kilojoules per mole, 0.0384 per cent of the internal energy figure, gives minus 3228.00 kilojoules per mole.
The Bigger jacket, smaller rise preset: the same gram of benzoic acid in 2500 g of water instead of 2000. The water level in the drawing has risen, the constant has gone to 12.092 kJ/°C and the rise has fallen to 2.1851 °C — while the heat card still reads qcal = 26.422 kJ and the per-gram card still reads -26.422 kJ/g over -3226.76 kJ/mol. A different apparatus, a different reading, the same answer.

Where to go next

A panel can put you in a state and let you move one thing at a time, and that is the end of what a panel is for. The bomb calorimeter equation guide is where the subject is written out: the four steps in order, the two corrections derived rather than quoted, and the problems worked from the thermometer reading to the molar answer.

Four posts own the pieces this lab takes as read. The first law of thermodynamics guide is where a rigid vessel doing no work comes from, the laws of thermodynamics guide puts that law in context, the ideal gas law guide owns the relation behind the gas-moles correction, and the pressure-volume diagram guide is where a constant-volume path enclosing no area is the subject.

Four more are worth having beside this one. The specific heat capacity guide and the specific heat capacity of water guide own the jacket term and the figure inside it, the latent heat guide owns the gross-and-net gap, and the difference between heat and temperature is the distinction the whole first card depends on. The energy conversion examples guide is where combustion sits among the rest.

On the tool side, the bomb calorimeter calculator takes typed figures and rearranges this chain in five directions, the reaction enthalpy calculator runs the step after it, the specific heat calculator does the jacket term on its own, the first law calculator does the law underneath all of it, and the water heating calculator is the quickest check on a jacket figure.

Four neighbouring pictures sit beside this one: the specific heat simulator, the latent heat simulator, the first law of thermodynamics simulator and the pressure-volume diagram simulator. Beyond them there is the simulation library, an index of calculators, and the blog for everything written out at length.

Frequently asked questions

Why does halving the sample halve the temperature rise but leave the kilojoules per gram alone?

Because the rise measures the burn and the per-gram figure measures the substance. Take benzoic acid from 1.00 g to 0.50 g and the rise card goes from 2.6422 to 1.3211 degrees Celsius and the heat card from 26.422 to 13.211 kJ, while the per-gram card holds at -26.422 kJ/g and the per-mole line at -3226.76 kJ/mol. That invariance is exactly what makes calibration worth doing.

What is the calorimeter constant, and why does this lab put it on two sliders?

It is the energy the whole assembly absorbs per degree, and it comes in two parts that no other free tool separates. The grid shows both: at the opening setting it reads 10.000 kJ per degree Celsius, of which 8.368 is the water jacket and 1.632 is the steel, the stirrer and the thermometer. Move either slider and the constant moves with it.

Why does the per-gram card carry a minus sign when the ladder marker sits at a positive value?

Because they are two different quantities with two different conventions. The card reports an energy change for the reaction, which loses energy, so it is negative and the sign is the physics rather than a formatting choice. The ladder plots a calorific value, which is conventionally quoted as a positive magnitude, and the grid labels it gross or net so the two cannot be confused.

Why does the enthalpy card say there is nothing to add when graphite is in the cup?

Because burning carbon consumes exactly as many moles of gas as it makes, so the correction term is exactly zero rather than merely small. The sub-line reads that the change in gaseous moles is 0 and the two cards are identical at -393.50 kJ/mol. It is decidable from the formula before any arithmetic starts, and the same is true of every carbohydrate.

Why does the grid call the figure gross, and what is the net value beside it?

Gross, because the bomb condenses the water its own reaction makes, so the energy given up on condensing is inside what the thermometer reads. Press Ethanol and the cell reads 29.7 gross, 26.7 net, with a gap of 10.10 per cent. Press Graphite and it reads 32.8 gross against 32.808 net, because carbon makes no water and the difference is rounding.

What is the smallest and the largest temperature rise these sliders can reach?

0.3926 degrees Celsius and 6.9168 degrees Celsius. The floor is 0.20 g of benzoic acid in 2500 g of water with the hardware at 3.000, and the ceiling is 1.50 g of propan-1-ol in 1500 g of water with the hardware at 1.000. Both were found by sweeping the real slider grids rather than by reasoning about corners, which matters because the smallest sample does not sit at the smallest rise.

Why does pressing Graphite while the stage says Calibrate move the stage to Measure?

Because a calibration run on graphite is not a calibration. Calibrate works by assuming the energy density of what is in the cup is already known, which is true of the standard calibrant and of nothing else here, so the stage forces benzoic acid into the cup. Choosing another sample is therefore a statement that you are measuring rather than calibrating, and the panel follows you.

Does the plausibility band prove my answer is right?

No, and one measured case shows why. The band checks a per-gram figure against four published bounds at 33.6, 48.64, 55.52 and 141.79 kJ/g, so it catches the impossible. A reader who measures the rise correctly but uses the water term alone as the constant gets 22.110 kJ/g against a true 26.422, which is 16.3 per cent low and still inside a plausible band.

What does this lab deliberately not draw?

Any picture of the reading climbing to its final value. The approach to a steady temperature is a separate topic with its own formula and its own page, so this panel draws the thermometer with the start and finish marked and the bracket between them, and no elapsed seconds anywhere. It also has four samples and no box to type an unknown one into.

References & formula source

  • Every figure on this page comes from one chain and five controls. The heat the assembly absorbs is the calorimeter constant multiplied by the temperature rise; the heat of the reaction is the negative of that; dividing by the sample mass gives kilojoules per gram and multiplying by the molar mass gives kilojoules per mole, which is an internal energy change because a rigid vessel does no pressure-volume work; and adding the change in gaseous moles multiplied by the gas constant and the reference temperature gives the enthalpy change. The calorimeter constant itself is the bomb hardware plus the mass of water multiplied by the specific heat capacity of water. The panel applies all of that; the written guide derives it and works the problems.
  • The constants are fixed once for the whole cluster and are printed exactly as they are used: the specific heat capacity of water is 4.184 joules per gram per degree Celsius, which is the thermochemical calorie by definition; the gas constant is 8.314 joules per mole per kelvin; and the reference temperature is 298.15 kelvin, which is the temperature all three fetched tables are quoted at. The gas constant is deliberately the four-figure value rather than the full SI one, so that every printed operation reproduces itself; the difference is 0.0056 per cent, which is twenty-two times smaller than the stated uncertainty on the calibrant.
  • The calibrant is the strongest provenance claim in this lab, because two independent sources agree. A government thermochemical compilation gives benzoic acid a molar enthalpy of combustion of -3228 kilojoules per mole with a stated uncertainty of 4, over a molar mass of 122.1213; an encyclopedia article quoting a calorie-era laboratory figure gives 6318 calories per gram. Converted, those are 26.4324 and 26.4345 kilojoules per gram, which agree to 0.0081 per cent. Benzoic acid is what the Calibrate stage forces into the cup for exactly that reason.
  • Benzoic acid has two legitimate energy densities and this cluster prints both, each with its label, because the 0.04 per cent between them is the payload sized on the calibrant itself. 26.422 kilojoules per gram is the constant-volume figure, which is what a bomb releases and what the reading cards carry. 26.43 kilojoules per gram is the constant-pressure figure, which is what the compilation tabulates and what the gross cell shows. Neither is ever printed on this site without saying which it is.
  • The reachable range of the temperature rise was swept rather than reasoned about, and the result corrected an earlier assumption in this cluster’s own planning notes. Over the real slider grids the smallest rise is 0.3926 degrees Celsius, at the smallest sample with the largest jacket and the largest hardware term, and the largest is 6.9168 degrees. The corner where all three sliders sit at their minimum gives 0.7263 degrees, which is a real and teachable state but is not the minimum, because the rise falls with both apparatus terms as well as with the sample mass.
  • The plausibility band is a filter and not a proof, and the page says so in the section on where this breaks down. It checks a per-gram figure against four published bounds: the highest oxygen-bearing value in the fetched set at 33.6 kilojoules per gram, the highest liquid hydrocarbon at 48.64, the highest non-hydrogen value anywhere in those tables at 55.52, and hydrogen at 141.79. A reader who measures the rise correctly and then uses the water term alone as the constant gets 22.110 kilojoules per gram against a true 26.422, which is 16.3 per cent low and still passes the band.
  • The result is a gross calorific value. Two independently fetched sources say so: one states that the higher heating value is what a bomb determines, including condensing any vapour produced, and the other explains that the vessel carries a small fixed amount of water so that its internal atmosphere is saturated and all the water the reaction makes ends up liquid. The gap between gross and net is the energy given up by that water, and the lab shows it directly in the graphite case, where no water is made and the two figures differ only by rounding.
  • Four external sources support the figures on this page and no others were used: a government thermochemical database entry for benzoic acid, with its uncertainty; an encyclopedia article on heats of combustion, for the gross-and-net distinction and for every value on the ladder; an encyclopedia article on calorimeters, for the apparatus, the split constant, the typical sample mass and jacket volume, the oxygen pressure and the three named corrections; and an encyclopedia article on enthalpy, for the definition the gas-moles step rests on. The one external link this page carries is appended automatically at the foot of this list.
  • A standard-compliant determination makes three corrections this lab does not: for the burning fuse wire, for nitric and sulfuric acid formed from nitrogen and sulfur in the sample and found by titrating the residual liquid, and for the electrical energy put in to start the ignition. They are codified in ASTM D5865 and ISO 1928 for solid fuels and in ASTM D240 and D4809 for liquid hydrocarbons. This page names them and attaches a numeric value to none of them, because none was fetched.
  • The simulation was built, gated and frozen before this page was written, and every string quoted here was read back out of the running file rather than computed here. A reference oracle at 193 assertions with none failing; a Node tier at 569 checks with none failing; a headless Chrome tier at 1717 checks with none failing across five widths; a compact-height pass measured at 947 pixels against a 1000 pixel ceiling; and a cross-browser pass in which the readouts came out identical in Chrome and in Firefox at each width tested.
  • What was not verified, stated as a gap rather than left to be assumed. Safari and WebKit: the build on the machine this was tested on cannot start, so Safari was never exercised and nothing here is evidence about it. Assistive technology: every slider carries a label and a spoken value, the drawing carries a full description and the preset status line is a live region, but no screen reader was driven and no real touch target was pressed. Canvas geometry below 150 pixels of drawing height is outside this lab’s supported range and is recorded as such in the build notes.
  • Nothing on this page is a measurement, and the lab is a forward model. It knows each sample’s tabulated energy density and works out the rise that would follow, so the Measure stage shows the arithmetic a student does rather than discovering an unknown, and no figure here has been checked against a physical experiment. The published heat of combustion for one of the four samples, propan-1-ol, is the fetched table value; the molar masses are derived from conventional atomic weights and validated against the one fetched molar mass. Verify any figure against your own source before you use it for anything that matters.
  • Further reading: Heat of combustion — Wikipedia